DE NOVO CLASSIFICATION REQUEST FOR

TEARSCIENCE, INC. LIPIFLOW® THERMAL PULSATION SYSTEM

REGULATORY INFORMATION

FDA identifies this generic type of device as:

Eyelid Thermal Pulsation System. An eyelid thermal pulsation system is an electrically-

powered device intended for use in the application of localized heat and pressure therapy to the eyelids. The device is used in adult patients with chronic cystic conditions of the eyelids, including meibomian gland dysfunction (MGD), also known as evaporative dry eye or lipid deficiency dry eye. The system consists of a component that is inserted around the eyelids and a component to control the application of heat and pressure to the eyelids.

NEW REGULATION NUMBER: 886.5200

CLASSIFICATION: II

PRODUCT CODE: ORZ

BACKGROUND

DEVICE NAME: LIPIFLOW® THERMAL PULSATION SYSTEM

SUBMISSION NUMBER: K093937 DATE OF DE NOVO: AUGUST 6, 2010

CONTACT: TEARSCIENCE, INC.

CHRISTY STEVENS, OD

PHONE: (919) 459-4815 FAX: (919) 467-3300

EMAIL: CSTEVENS@TEARSCIENCE.COM

REQUESTER’S RECOMMENDED CLASSIFICATION: CLASS II

INDICATIONS FOR USE (IFU)

The LipiFlow® System is intended for the application of localized heat and pressure therapy in adult patients with chronic cystic conditions of the eyelids, including meibomian gland dysfunction (MGD), also known as evaporative dry eye or lipid deficiency dry eye.

LIMITATIONS

1. Caution: Federal Law restricts this device to sale by or on the order of a physician.

De Novo Summary (K093937) Page 1 of 8

PLEASE REFER TO THE LABELING FOR A MORE COMPLETE LIST OF

WARNINGS, PRECAUTIONS AND CONTRAINDICATIONS.

DEVICE DESCRIPTION

The physician uses the LipiFlow® System in an in-office procedure to control the application of warmth and massage to the eyelids. Two components comprise the LipiFlow® System, the Disposable unit and the Handheld Control System (HCS).

The Disposable unit is a sterile, single-use, biocompatible unit that is inserted around the patient’s eyelids. The Disposable unit consists of a combined Eye Cup and Lid Warmer with attached tubing and wiring that connect to the Control unit with a connector. The Eye Cup contacts the outer eyelid and contains a soft, flexible bladder that intermittently inflates with air to provide controlled massage pressure to the eyelids. The Lid Warmer contacts the inner eyelid surface and provides controlled, outward directional (away from the eye and towards the eyelid) heat to the inner eyelid. The Lid Warmer has a smooth surface and edges, where the circumference rests lightly on the conjunctiva of the eye. The Lid Warmer shape vaults above the eye surface to prevent corneal contact. The Lid Warmer has an integrated insulator (also referred to as the Insulating Scleral Lens) to shield the eye from thermal transfer and redundant temperature sensors to ensure precise control of the temperature.

The HCS is a battery-operated, hardware interface that allows the physician to control the application of heat and pressure, which is delivered via a single, 12-minute treatment. The HCS regulates the level of heat and pressure that can be applied during treatment. The HCS consists of a visual Light Emitting Diode (LED) display, power switch, push button, electronic circuit board, pump, pressure regulator, pressure sensor, dump valve, connector and battery compartment. The HCS displays information to the physician including: the temperature at the Disposable; line air-pressure to the Disposable; treatment time; low pressure selection indicator;

and system messages, such as low battery, a system error, or inadequate connection to the Disposable. By visualizing the temperature and pressure LED displays, the physician can determine when the therapeutic temperature is reached and the relative pressure being applied.

The push button starts the application of heat and pressure and can be depressed again at any time to pause the treatment, which immediately stops the therapeutic application. In the event that the physician desires to reduce pressure for patient comfort, the pressure can be reduced by 30% (Low Pressure Mode) by pressing and holding the push button. The HCS uses three AA (1.5V) Lithium disposable batteries.

–  –  –

The Disposable consists of a combined Eye Cup and Lid Warmer, which connect to the HCS:

In this cross-sectional representation, the Disposable positions around the eyelids during device

use:

–  –  –

BIOCOMPATIBILITY/MATERIALS

The patient-contacting materials including the adhesive in the Disposable of the LipFlow System were evaluated for cytotoxicity (ISO 10993-5, Biological evaluation of medical devices-Part 5: Tests for in vitro cytotoxicity), irritation and sensitization (ISO 10993-10, Biological evaluation of medical devices-Part 10: Tests for irritation and delay-type hypersensitivity) in accordance with the recommendations in ISO 10993-1, Biological evaluation of medical devices - Part 1: Evaluation and testing, and in FDA 510(k) Memorandum 95-1 for materials with limited mucosal membrane contact. Additionally, the sponsor also provided a certificate of compliance from a test laboratory certifying that the adhesive passed the following tests; irritation (ISO 10993-10, Biological evaluation of medical devices-Part 10: Tests for irritation and delay-type hypersensitivity), acute systemic toxicity (ISO 10993-11, Biological evaluation of medical devices-Part 11: Tests for systemic toxicity), cytotoxicity (ISO 10993-5, Biological evaluation of medical devices-Part 5: Tests for in vitro cytotoxicity), hemocompatibility (ISO 10993-4, Biological evaluation of medical devices-Part 4: Selection of tests for interactions with blood), implantation (ISO 10993-6, Biological evaluation of medical devices-Part 6: Tests for local effects after implantation) as well as the USP physiochemical test. Material Safety Data Sheets (MSDSs) on the materials were also provided and they do not raise any concerns that have not been addressed by the testing performed by the sponsor.





The materials used in the Disposable are, therefore, believed to be biocompatible when the Disposable is used as intended.

–  –  –

The proposed shelf life of 6 months has been validated using accelerated aging. The

accelerated aging tests include:

• Tensile strength testing - ASTM F88:2007, “Standard test method for seal strength of flexible barrier materials”

• Burst /creep pressure testing - ASTM F1140:2007, “Standard test methods for internal pressurization failure resistance of unrestrained packages”

• Dye penetration testing – ASTMF 1929-98:2004, “Detecting seal leaks in porous medical packaging by dye penetration”

• Bubble test - ASTM F2096:2004, “Standard test method for detecting fross leaks in medical packaging by internal pressurization”

–  –  –

ELECTRICAL, MECHANICAL AND THERMAL SAFETY

During the Electrical Engineering/ Hermeticity review of K093937, an Engineering Analysis was conducted of the hermetic sealing of the device with regard to device performance and safety. The chief concern was that if the device exhibited a leak, fluids could enter the device and alter the thermal sensing and thermal transfer characteristics.

The reviewer requested validation studies of the integrity of the glue joint forming the hermetic barrier, as well data on the potential harmful effects that could arise if moisture were to penetrate this barrier. The company responded to these requests and demonstrated that the proposed testing would detect a device with a compromised seal. The company further demonstrated that in the event that a seal was compromised, the safety of the device would not be affected. Deficiencies which were subsequently addressed to the satisfaction of the reviewer related to the integrity of the hermetic seals, validation testing for the vacuum testing of the glue joint, information on the effect of saline-filled tears on the electrical performance of the heating element, and changes in the heat transfer properties if fluid fills the air gap between the heater and the insulator shell.

A previous Electromechanical Engineering review of this device indicated deficiencies relating to the Electrical Inspection Criteria, the Insulator Assembly Inspection Criteria, the Quality Control Procedure, and the Protocol Verification results including verification of an applied pressure (b)(4) mm Hg. In addition, no test results had been provided for the Controller Board Incoming Inspection Criteria or the Final Inspection Procedure. The reviewer notes that the sponsor has satisfactorily addressed each of the concerns. The

device is considered to adhere to the following electrical safety standards:

–  –  –

De Novo Summary (K093937) Page 5 of 8 maximum corneal temperature measured immediately after device treatment using a thermal camera during performance testing with human eyes in vivo reached (b)(4) ° C.

Finite Element Analysis modeling device usage in worst case conditions of no blood flow, and thin eyelids resulted in a maximum predicted corneal temperature of (b)(4) ° C. The safeguards include a current limiting switch, redundant temperature regulators, and temperature shutoff circuitry monitoring redundant temperature sensors.

–  –  –

SUMMARY OF CLINICAL INFORMATION

A summary of the clinical study which was conducted follows:

The study objective was to evaluate the clinical utility, safety and effectiveness of the LipiFlow® System compared to warm compress therapy using a commercially available chemical heat device (iHeatTM Portable Warm Compress System) for application of localized heat therapy in adult patients with chronic cystic conditions of the eyelids, including meibomian gland deficiency (MGD), also known as evaporative dry eye or lipid deficiency dry eye.

A total of 139 subjects (278 eyes) were enrolled at nine sites with randomization of 69 subjects (138 eyes) to the LipiFlow® group and 70 subjects (140 eyes) to the Warm Compress Control group. LipiFlow® subjects received a single, 12-minute in-office treatment with the LipiFlow® System at the Treatment visit. LipiFlow® subjects were followed at 1 day, 2 weeks and 4 weeks after treatment. Control subjects received the initial 5-minute iHeatTM System therapy per the device labeling at the Treatment visit.

Control subjects were instructed to use the warm compress therapy for 5 minutes daily at home until the 2-week visit. After 2 weeks the Control group crossed over to receiving a single LipiFlow® treatment.

Effectiveness parameters were Meibomian Gland Assessment, Tear Break-up Time and Standard Patient Evaluation of Eye Dryness (SPEED) and Ocular Surface Disease Index (OSDI) Dry Eye Questionnaires.

Safety parameters were Discomfort/Pain Evaluation, Ocular Surface Staining, Intraocular Pressure (IOP) by Goldmann tonometry, Slit Lamp and Dilated Retinal Exam, and Best Spectacle Corrected Visual Acuity (BSCVA). Safety parameters were used to assess for Adverse Events.

The LipiFlow® System met the primary study effectiveness endpoint with a statistically significant (p0.0001) greater improvement at 2 weeks from baseline in the average number of meibomian glands yielding clear liquid secretion as compared to the Warm Compress Control. The LipiFlow® group showed a mean improvement in tear break-up time and a mean reduction in dry eye symptoms at 2 weeks from baseline with an effect greater than the Warm Compress Control. A single 12-minute treatment with the

–  –  –

No device-related serious adverse events or unanticipated adverse device effects were reported in the study. Four eyes (2.9%) in the LipiFlow® group had device-related nonserious adverse events, including three eyes with moderate eyelid pain and one eye with moderate conjunctival vascular injection (redness). All of these adverse events resolved during the 4-week study without sequelae or medical treatment.

In conclusion, the study results demonstrate that the benefit of the LipiFlow® System in providing heat and pressure therapy for patients with meibomian gland dysfunction and dry eye symptoms outweighs the risk.

A previous Clinical review of this device indicated deficiencies relating to the standard of care for meibomian gland deficiency, as well as the lack of data to support two of the Indications for Use originally sought by the sponsor. Subsequently, each of the deficiencies has been satisfactorily addressed, as acceptable clarification was provided regarding the standard of care, and the unsupported Indications were removed. The clinical data, combined with the sponsor responses to the deficiencies, support the conclusion that the test device/ treatment is both safe and effective, compared to the control device/ treatment.

As part of the analysis for the primary safety endpoint, the significance of the observed device-related adverse events rates was determined. The clinician concluded the observed difference in device-related adverse events was not clinically significant between the two groups.

LABELING

The labeling for the LipiFlow® Thermal Pulsation System is consistent with the clinical data and covers all the hazards and other clinically relevant information that may impact safe and effective use of the device. The labeling is sufficient and satisfies the requirements of 21 CFR Part 801.109 Prescription devices.

RISKS TO HEALTH



Pages:   || 2 |


Similar works:

«GCSE Options 2016 1 Contents AN INTRODUCTION TO GCSEs FROM SEPTEMBER 2016 MAKING CHOICES AND DECISIONS MAKING CHOICES AND DECISIONS ENGLISH LANGUAGE (AQA) ENGLISH LITERATURE (AQA) MATHEMATICS (EDEXCEL) BIOLOGY (EDEXCEL iGCSE) CHEMISTRY (Edexcel iGCSE) PHYSICS (Edexcel iGCSE) ART AND DESIGN (OCR) CLASSICAL CIVILISATION (OCR) DRAMA (EDEXCEL) MODERN FOREIGN LANGUAGES – French, German, Spanish (AQA).15 GEOGRAPHY (EDEXCEL B) HISTORY (EDEXCEL) HOME ECONOMICS (FOOD AND NUTRITION) (OCR) ICT GCSE...»

«Summer 2012 TIAA-CREF Asset Management Shopping: a whole new meaning Martha Peyton, Ph.D., Managing Director TIAA-CREF Global Real Estate, Strategy & Research Savvy investors in the retail sector are well aware of the necessity of keeping up with change. in fashion, in customer behavior, in geography. and so on. Every decade since the era of the general store on Main Street has seen the retail sector rattled by new products and new ways to sell them. Focusing just on the latter, the...»

«30029-XX Entocort® EC (budesonide) Capsules Rx only DESCRIPTION Budesonide, the active ingredient of ENTOCORT® EC capsules, is a synthetic corticosteroid. It is designated chemically as (RS)­ 11β, 16α, 17,21-tetrahydroxypregna-1,4-diene-3,20-dione cyclic 16,17-acetal with butyraldehyde. Budesonide is provided as a mixture of two epimers (22R and 22S). The empirical formula of budesonide is C25H34O6 and its molecular weight is 430.5. Its structural formula is: C H 2O H C O CH3 H O HO C O...»

«TOP TIPS T O H E L P YO U RIASE MORE VISIT EVERYDAYHEROCOMMUNITY.COM/ OR CALL 1300 798 768 TO LEARN MORE. EVERYDAY HERO • 1300 798 768 • EVERYDAYHERO.COM.AU PROPELLING PASSION FOR CHANGE CONTENTS T IPS F O R RA I S I N G M ON EY FO R THE CAUSE YO U CA RE A B O UT 3 RAIS ING M ON EY AS A T E A M 4 G IVING TH I N G S U P TO RA I SE MO N EY FO R YO UR CAUSE 5 RAIS ING M ON EY U SI N G FACEB O O K 6 RAIS ING M ON EY U SI N G T W ITTER 7 RAIS ING M ON EY U SI N G I N STAGRA M 8 TO P WAYS TO RA I...»

«Many of the following comments are taken from a sermon given by Dr. Ed DeVires on August 10, 1998 at King James Baptist Church in Arlington, Texas. Some editing has been done for clarity purposes. We have endeavored to insure that the original meaning of the sermon remains intact. All scriptural references’ are from the King James version of the Bible. Let us begin with a basic scripture which teaches us to glorify God in our bodies, as well as our spirits. I COR 6:19-20 What? know ye not...»

«The Bosnian State a Decade after Dayton SUMANTRA BOSE A decade on from the Dayton peace settlement, this essay sets out to examine two questions. First, is the consociational and confederal paradigm established by the Dayton agreement, and subsequently institutionalized, the appropriate framework for the Bosnian state? It will be suggested that in the circumstances that prevail, this framework does in fact provide the most feasible and most democratic form of government for Bosnia’s...»

«i HP Creative Scrapbook Assistant Software for quick and easy printable layouts and unique page elements User’s Guide ii Contents Introduction.................................. 4 About this User’s Guide..................... 5 Visit the HP Scrapbooking Web site............. 6 Getting started........................... 6 1 Installing and starting the program................. 7...»

«© 2010 The two sisters build the Twelve Tribes of Israel, and bear a positive influence on each other. by Dina Coopersmith When we last left off, Rachel was learning a tough lesson in the realities of life as a matriarch: Why were the matriarchs barren? Because God desires the prayers and conversations of the righteous. He said: “They are beautiful, they are rich; if I give them children, when will I ever hear from them? (Midrash Tanchuma Toldot 9) It's not easy being a matriarch! Of the...»

«Tantasavasdi, C., Srebric, J., and Chen, Q. 2001. Natural ventilation design for houses in Thailand, Energy and Buildings, 33(8), 815-824. Natural Ventilation Design for Houses in Thailand Chalermwat Tantasavasdia, Jelena Srebricb, Qingyan Chenc* a Architecture Program, Thammasat University, Room 713 Faculty of Engineering Building, Rangsit Campus, Pathumthani 12121, Thailand b Department of Architectural Engineering, The Pennsylvania State University, 222 Engineering Unit A, University Park,...»

«UPSET TRENDS IN FLIP-FLOP DESIGNS AT DEEP SUBMICRON TECHNOLOGIES By Vijay Benakanakere Sheshadri Thesis Submitted to the Faculty of the Graduate School of Vanderbilt University in partial fulfillment of the requirements for the degree of MASTER OF SCIENCE in Electrical Engineering December, 2010 Nashville, Tennessee Approved: Professor Bharat L Bhuva Professor Robert Reed ACKNOWLEDGEMENTS I would like to thank my advisor Professor Bharat Bhuva, Professors Robert Reed and Ron Schrimpf for their...»

«Haldex Automates Lean Manufacturing Processes and Streamlines Operations with Microsoft Dynamics AX The phased implementation of Microsoft Dynamics AX and eBECS’ Lean Automotive for Dynamics AX supports Haldex’s vision and fundamental commitment to achieving world class productivity while developing their employees’ competence and creativity. Haldex has already implemented Microsoft Dynamics AX in operations across the USA and Canada, and is » continuing the global rollout through 2006....»

«1 Getting Started with Exodus By Lauren Stouffer and Ted Hildebrandt Copyright © 2012 Introduction to Exodus Exodus, the second book in the Torah, begins where Genesis ended—with the children of Israel living in Egypt. The original Hebrew title of the book “and these are the names,” is taken from the first two words of the book in the Hebrew. The title “Exodus” (“the way out”) is taken from the Greek Septuagint, or LXX1 explicating the content of the book as being about the way...»





 
<<  HOME   |    CONTACTS
2016 www.dissertation.xlibx.info - Dissertations, online materials

Materials of this site are available for review, all rights belong to their respective owners.
If you do not agree with the fact that your material is placed on this site, please, email us, we will within 1-2 business days delete him.