www.thinkSRS.com

Pharmacopeia vs. Thermodynamic

Melting Point Determinations

Application Note #3

Pharmacopoeia Melting Point

The procedural rules for melting point determinations are defined in the

pharmacopeias. The medical handbooks include minimum requirements for the

design of the melting point apparatus and for performing the measurements.

Automated melting point determination procedures are generally included. Very

often, the pharmacopeias also list special methods for difficult or unusual cases of melting point determination.

Figure 1. Standard capillary tubes used in melting point determinations.

The pharmacopeias regard the capillary technique as the standard method for melting point determination. In this method, thin glass capillary tubes containing packed samples of the substance are introduced into a heating stand that is continuously being heated up.

The capillary technique is the standard method used for melting point determinations in most organic and pharmaceutical chemistry laboratories.

The procedures for melting point determinations in the pharmacopeias call for a fixed rate of temperature rise in the “heating stand” (i.e. liquid bath and/or metal block), typically between 0.2 °C/min and 2 °C/min, within a temperature range that brackets the expected melting temperature of the compound. The oven continues to heat at a constant rate until the sample is completely melted (i.e.

clear point). In determining the melting point, the temperature on the thermometer at the clear point (and/or sometimes at the meniscus point) is recorded. This is not the temperature of the sample itself but rather that of the Stanford Research Systems Phone: (408) 744-9040 Pharmacopeia vs. Thermodynamic MP Determinations www.thinkSRS.com heating stand where the thermometer is located (i.e. the temperature is determined by the heating medium).

When different heating rates are used for determining the melting point of a sample, the values obtained for the clear point are dependent on the temperature ramping rate. Because the heat transfer from the heating stand to the sample cannot be increased proportionally to the temperature ramping rate, the temperature in the heating stand rises to a higher level with faster rates of heat increase than it does with lower rates. As a result, the faster the temperature is ramped, the higher the clear points are found to be. Due to this dependence on heating rate, measurements taken for melting points are comparable with one another only if they were taken using the same ramping rates. Any variance from the temperature ramping rates specified by pharmacopoeia procedures must be properly recorded for Good Laboratory Practice documentation.

Table 1 is an example of the dependence of the clear point determination on ramping rate for phenacetin samples (MP: 133 ºC) analyzed with an OptiMelt system. As expected, the clear point increases with increasing ramping rates.

The need to specify ramping rates along with the results of the melt is demonstrated by this data.

–  –  –

The capillary techniques described in early pharmacopeias were mostly concerned with visual detection of the melting point. Recent reviews of the published procedures have been supplemented to account for automated detection of the melting point, such as the Digital Image Processing technology used by the OptiMelt system.

The heating stands used by commercial melting point apparatuses are divided

into two categories:

• Liquid BathThe capillaries are immersed in a liquid bath (typ. silicone oil) that is continuously being heated up.

• Metal BlockA small, dry thermal block (metal oven) has proven to be a good alternative to a liquid bath.

–  –  –

The melting point procedures described in early pharmacopeias were designed for “liquid bath” heating stands. However, in recent years most monographs have been updated and supplemented to include the more prevalent metal block setups.

Tip A slight drift in MP results is expected when the MP numbers obtained with a liquid bath oven are compared to those obtained with a metal block stand. This difference is well known and generally ignored since it typically falls within the intrinsic uncertainty of the measurement.

Thermodynamic Melting Point The transition from solid to liquid does not take place instantaneouslyit requires a finite amount of time. The melting process begins at the point where the first particles of the bulk substance turn into the liquid statethe onset point. The end of the melt is reached when the last solid particles have gone over into the liquid phasethe clear point. During the entire melting process of a pure compound, the temperature of the pure substance remains constant (thermodynamic melting point) while heat is transferred from the heating stand to the sample. The heating stand itself experiences a range of temperatures that depends on the selected heating rate.

When determining the melting point according to pharmacopoeia, the temperature of the heating stand at the end of the melt (i.e. clear point) is read.

That single temperature record depends on the temperature ramping rate. It ignores the range between the start and the end of the melt, and it is not the “true” thermodynamic melting point of the pure compound.





–  –  –

Figure 2. Graphical representation of the thermodynamic correction Figure 2 is a simple representation of the sequence of events that take place during the melting of a pure substance.

At the start of the melting (time = t0), the block and the sample are at approximately the same temperature. As soon as the melt starts the sample temperature stabilizes while the block continues heating up. As the melt progresses, the sample remains at constant temperature (thermodynamic melting point, MPtherm) while the block continues to heat up. Heat is constantly transferred from the block to the sample at a rate that is proportional to the temperature gradient between the sample and the block. The temperature of the block at the end of the melt (time = tf) is recorded as the clear point (pharmacopoeia melting point, MPpharma). The thermodynamic correction is

defined as:

–  –  –

and is represented as a function of the ramping rate, r.

In order to obtain the thermodynamic melting temperature of a pure substance, it is necessary to calculate and subtract the thermodynamic correction from the detected clear point. This calculates back to the temperature at the beginning of the melt so that the value obtained includes practically no dependence on the temperature ramping rate used for the determination.

Stanford Research Systems Phone: (408) 744-9040 Pharmacopeia vs. Thermodynamic MP Determinations www.thinkSRS.com Instruments with automated melting point determination facilities often pack enough data analysis infrastructure to automate the thermodynamic correction procedure: (1) the clear point is accurately identified and recorded, (2) the ramping rate is known and carefully controlled, and (3) the thermodynamic correction algorithm can be programmed and stored in memory. Knowledge of the functional dependence of the thermodynamic correction, ∆T( r ), on r is the only requirement to automate the thermodynamic correction process. A parametric derivation of that functional dependence is presented in the following section.

Thermodynamic Correction At any given time, t, during the melt, the amount of heat, dQ(t), transferred from

the heating stand to the sample during a time dt is:

–  –  –

where, T: temperature of the heating stand, [°C] t: is the time variable, [min] r = dT/dt: temperature ramping rate of the instrument [°C/min] α: heat transfer constant for the melting point apparatus [calories/(°C. min)]

Substituting dt with dT in eqn. 2, leads to:

–  –  –

Integration of the heat transferred from the block to the sample, over the entire melting process, provides the “heat of fusion” of the sample, ∆Hf [calories], which is dependent on its mass but independent of the ramping rate, r.

–  –  –

Calculation of the integral term leads to the analytical expression:

.

∆Hf = [0.5 α/r]. (MPpharma − MPtherm)2 (eqn. 5) which can be rearranged to provide an equation for the thermodynamic

correction:

∆T (r)= MPpharma − MPtherm = [2. ∆Hf / α]1/2. r1/2 = (Thermo CF). r1/2 (eqn. 6)

–  –  –

According to eqn. 6, the thermodynamic correction is directly proportional to the square root of the ramping rate, and in order to calculate the thermodynamic melting temperature, it is necessary to know the value of the Thermodynamic

Correction Factor (Thermo CF) for the melting point apparatus:

MPtherm = MPpharma (clear point) – (Thermo CF). r 1/2 (eqn. 7)

The above theory indicates that the value of the thermodynamic correction factor (Thermo CF), depends, among other things, on the (1) heat of fusion of the sample, (2) the thermal conductivity of the sample, (3) the thermal conductivity of the glass capillary, (4) the sample preparation/packing method, and (5) the geometry/construction of the oven. In practice, the value of Thermo CF is both compound and instrument dependent and must be calculated through empirical determination.

Thermodynamic Correction with OptiMelt A user-programmable Thermodynamic Correction Factor (Thermo CF) can be programmed into OptiMelt and used by the instrument to automatically calculate Thermodynamic Corrections, ∆T (r) = (Thermo CF). r1/2, at the end of a meltpoint determination.

Tip

• Thermodynamic corrections are appended to melt-point reports if Thermo MP is set to “Calculate” in the Options menu of the Melt Screen.

• Use the ThermoCF menu button to program the value of Thermo CF for the compound being measured.

• Compound specific thermodynamic correction factors can be stored within the analysis methods saved to memory.

• Use the thermodynamic corrections calculated by your OptiMelt to correct “ramp-dependent” clear points (manual or automatic) and to obtain a true thermodynamic melting point for your samplesindependent of the rate of heat rise used.

• Empirical measurements have shown that in most cases a good approximation to the Thermodynamic Correction Factor (Thermo CF) for OptiMelt is a value of ≈1.0, when r is expressed in standard units of [°C/min].

• Calculation of your sample’s Thermodynamic Correction Factor (Thermo CF) is recommended when accuracy is required. See next section for details.

–  –  –

Calculation of Thermodynamic Correction Factor The Thermodynamic Correction Factor (Thermo CF) is compound specific and

depends on several factors:

• Specific heat of fusion of the sample

• Mass of the sample

• Thermal conductivity of the sample

• Thermal conductivity of the glass capillary

• Sample preparation method

• Geometry of the oven Experimental measurements have shown that in most cases a good approximation to the Thermodynamic Correction Factor (Thermo CF) for OptiMelt is a value of ≈1.0, when r is expressed in standard units of [°C/min]. However, empirical calculation of the Thermodynamic Correction Factor is recommended when more accurate results are required for a specific compound.

A Step-by-step calculation procedure is described below.

Step 1. Perform complete pharmacopoeia melting point determinations on your sample at six different temperature ramping rates: 0.

1, 0.2, 0.5, 1, 2 and 5°C/min.

• Three samples are analyzed simultaneously for each ramping rate. Load three capillaries with a 2 to 3 mm column of sample and insert the tubes into the three adjacent sample slots of the OptiMelt oven.

• To increase the accuracy of results, load the capillaries using the same grinding, loading and packing techniques used for routine MP determinations.

• For each ramping rate, average the detected “clear points” (according to pharmacopoeia) for the three samples to obtain the final value.

• To increase the accuracy of results, use manual (automatic) clear point readings if manual (automatic) determinations are routinely performed in the lab.

Step 2. Plot the resulting Clear Point records versus the square root of their corresponding temperature ramping rate (MPpharma vs r1/2).

A linear dependence between the two sets of numbers should be observed. Calculate the slope of the straight line, and program the result into the Thermodynamic Correction Factor (Thermo CF) of your OptiMelt.

Step 3. Use the Thermodynamic Correction Factor to calculate the thermodynamic melting point of the sample from the measured clear points (eqn.

–  –  –

7). An agreement within the accuracy of measurement should be observed for all calculated thermodynamic melting points.

An example of this calculation procedure is included below.

Thermo CF Calculation Example In order to demonstrate the simplicity of the calculation procedure described in the previous section, a series of melts were performed on a phenacetin sample with ramping rates 0.1, 0.2, 0.5, 1.0, 2.0 and 5.0 °C/min.

Figure 3 summarizes the results and shows the expected linear relationship between the clear point determinations and the square root of the ramping rate.

The slope of the straight line corresponds to a thermodynamic correction factor, Thermo CF= 1.9 for phenacetin samples.

–  –  –

Figure 3. “Clear point temperature vs.

square root of ramping rate” for a phenacetin sample melted at 0.1, 0.2, 0.5, 1.0, 2.0 and 5.0 ˚C/min. The slope of the straight line, Thermo CF=1.9, is programmed into OptiMelt as the thermodynamic correction factor (Thermo CF) for this compound.



Pages:   || 2 |


Similar works:

«MSHQ Podcast Session #16: Interview with Mount Sinai – All About FlexMed show notes at: http://www.medicalschoolhq.net/session16 Intro Announcement: You’re listening to the Medical School HQ Podcast, online at medicalschoolhq.net, session #16! Ryan Gray: Hello and welcome back to another awesome session here at the Medical School HQ podcast. I am your host Ryan Gray, and we are THE podcast about medical school. From the pre-med process through residency, we hope to take your knowledge of...»

«RESNA Position on the Application of Seat-Elevating Devices for Wheelchair Users Rehabilitation Engineering & Assistive Technology Society of North America 1700 N. Moore Street, Suite 1540 Arlington, VA 22209 Phone: 703-524-6686 Fax: 703-524-6630 Approved by RESNA Board of Directors September 2005 RESNA Position on the Application of Seat-Elevating Devices for Wheelchair Users RESNA Position on the Application of Seat-Elevating Devices for Wheelchair Users Wheelchair mobility is often only...»

«10 May 2006 WHO Position Paper on tetanus vaccines: Selected references Epidemiology Bytchenko B. Geographical distribution of tetanus in the world. Bull World Health Organization 34: 71-104, 1966. (Abstract not available). Galazka A, Gasse F. The present status of tetanus and tetanus vaccination. Curr Top Microbiol Immunol. 1995;195:31-53. (Abstract not available). Pedalino B, Cotter B, Ciofi degli Atti M, Mandolini D, Parroccini S, Salmaso S. Epidemiology of tetanus in Italy in years...»

«TMD This brochure is produced by the American Academy of Orofacial Pain The American Academy of Orofacial Pain is an organization of health care professionals dedicated to alleviating pain and suffering through the promotion of excellence in education, research and patient care in the field of Orofacial Pain and associated disorders. This brochure is intended to provide general information on temporomandibular disorders and is not a substitute for careful evaluation by a physician or orofacial...»

«Do Unemployment Benefit Extensions Explain the Emergence of Jobless Recoveries?⇤ Kurt Mitman† Stanislav Rabinovich‡ December 27, 2014 Abstract In this paper, we show that countercyclical unemployment benefit extensions can lead to jobless recoveries. We augment the standard Mortensen-Pissarides model to incorporate unemployment benefit expiration and state-dependent extensions of unemployment benefits. In the model, an extension of unemployment benefits slows down the recovery of...»

«Registration of Medicines Stability SADC GUIDELINE FOR STABILITY TESTING This guideline is intended to provide requirements to applicants wishing to submit applications for the registration of medicines. It represents SADC current thinking on the safety, quality and efficacy of medicines. It is not intended as an exclusive approach. Member States reserves the right to request any additional information to establish the safety, quality and efficacy of a medicine in keeping with the knowledge...»

«The CPS Response to Child Neglect An Administrator’s Guide to Theory, Policy, Program Design and Case Practice Editors Thomas D. Morton Barry Salovitz Published by National Resource Center on Child Maltreatment Operated by ACTION & For Child Protection A Service of the Children’s Bureau, U.S. Department of Health and Human Services © August 2001 The CPS Response to Child Neglect An Administrator’s Guide to Theory, Policy, Program Design and Case Practice Editors Thomas D. Morton Barry...»

«
 1
 Memory for pictures: Sometimes a picture is not worth a single word Joyce M. Oatesa and Lynne M. Rederb a American University b Carnegie Mellon University Mailing Address: Department of Psychology Carnegie Mellon University Pittsburgh, PA 15213 e-mail: jo5436a@student.american.edu or reder@cmu.edu Preparation of this chapter was supported by grant R01-MH52808 from the National Institute of Mental Health to the second author. Words: 5,586 
 2
 Memory for pictures: Sometimes a...»

«TMJ Implants A Consumer Information Update April 2001 Dear Reader: This is an updated introductory letter from the current 1999 TMJ implant consumer handbook entitled, TMJ Implants – A Consumer Informational Update 1999. While the Food and Drug Administration (FDA) is in the process of updating the handbook, we want to provide you with some new updates and the current information.Background: The FDA’s regulation of new medical devices entering the market began with the 1976 Medical Device...»

«Pathology in Australia Pathology: the Profession Pathology is literally the study of disease and is the science upon which medicine is based. Pathologists are medical specialists who must undergo a minimum of 11 years medical training to qualify. There are six areas of specialty within pathology, viz Anatomical Pathology, Haematology, Chemical Pathology, Microbiology, Immunopathology, and Genetics; this diversity in expertise throughout private and public sectors in all states is essential to...»

«CONSERVATION AND RECOVERY STRATEGY FOR SAGE-GROUSE (Centrocercus urophasianus) AND SAGEBRUSH ECOSYSTEMS WITHIN THE DEVIL’S GARDEN / CLEAR LAKE POPULATION MANAGEMENT UNIT Clear Lake Sage-Grouse Working Group Compiled By Marc R. Horney, Ph.D., CRM Rangeland Management Specialist Klamath Basin Watershed Team USDA-NRCS June 11, 2008 TABLE OF CONTENTS Chapter 1: General Overview of Sage-Grouse Status As A Protected Species Biology Taxonomy and Description Distribution Life History and Habitat...»

«CURRICULUM VITAE AND RESEARCH WORK of CHRISTOS P. LOIZOU http://www.medinfo.cs.ucy.ac.cy/ January 2012 Resumé of Christos P. Loizou Executive Summary I. General Information · M.Sc. (Dipl.-Ing) in Electrical Eng. and Telecommunications, Univ. of Kaisserslautern, Germany, 1990 · Ph.D. in Computer Science, Kingston Univ., UK, Ultrasound image analysis of the carotid artery, 2005 · Research interests: medical imaging, signal and image processing, and health Telematics · Assistant at the...»





 
<<  HOME   |    CONTACTS
2016 www.dissertation.xlibx.info - Dissertations, online materials

Materials of this site are available for review, all rights belong to their respective owners.
If you do not agree with the fact that your material is placed on this site, please, email us, we will within 1-2 business days delete him.